For each of the nine reference points on the screen, the distance between the detected pressed position and the nearest reference point is displayed. The distance is displayed continually near each reference point.
If all the distances between each detected pressed position and the nearest of the five reference points on the screen (at the four corners and the center) are within ±12px, the [OK] button appears. (However, even when the [OK] button is being displayed, if you press the screen again and the distance exceeds ±12px, the [OK] button disappears.)
If you press the [OK] button, the inspection implementation status switches to “Adjusted” and the screen returns to the self-inspection menu.
If you press the [Back] key, the inspection implementation status does not switch and the screen returns to the self-inspection menu.